facebook

RION VM-82A General-Purpose Vibration Meter

 

Key Features

  •  Operation panel with optimized button layout makes mode switching and setup easy and fast
  •  Wide range of measurement applications supported by selecting different accelerometers
  •  Backup function instantly reactivates previous settings at next power-on
  •  Convenient USB interface allows transfer of saved data to a computer
  • Compact dimensions and light weight: only 270 grams including batteries

SPECIFICATIONS RION VM-82A: GENERAL-PURPOSE VIBRATION METER

— Piezoelectric Accelerometer PV-57エ (supplied)

Type: Shear-type piezoelectric accelerometer (CCLD compatible)

Sensitivity: 5.1 mV/(m/s2) (±15 %) 80 Hz, 23 °C

Frequency range: 1 Hz to 5 kHz (±10 %)

Dimensions / Weight: 17 (width across hexagonal flat) × 49 mm / 45 g

— Applicable standards: CE marking, WEEE Directive, Chinese RoHS. EMC standards: IEC 61326-1、CISPR 11、IEC 61000-6-2

— Measurement range (using PV-57エ)

ACC (Acceleration): 0.02 to 200 m/s2 EQ PEAK 1 Hz to 5 kHz

VEL (Velocity): 0.3 to 1 000 mm/s RMS 3 Hz to 1 kHz; 0.1 to 1 000 mm/s RMS 10 Hz to 1 kHz

DISP (Displacement): 0.02 to 100 mm EQ PEAK 3 Hz to 500 Hz; 0.001 to 100 mm EQ PEAK 10 Hz to 500 Hz

— Frequency range

ACC (Acceleration): 3 Hz to 1 kHz, 3 Hz to 5 kHz, 1 Hz to 100 Hz, 3 Hz to 20 kHz

VEL (Velocity): 10 Hz to 1 kHz, 3 Hz to 1 kHz

DISP (Displacement): 10 Hz to 500 Hz, 3 Hz to 500 Hz

Values represent the range measured to about 10 % attenuation from flat response, due to high-pass filter or low-pass filter action. Electrical characteristics for velocity from 10 Hz to 1 kHz are compliant with the frequency range requirements of JIS B 0907 “Mechanical vibration of rotating and reciprocating machinery – Requirements for instruments for measuring vibration severity”.

— Measurement full-scale range

For accelerometer PV-57エ and accelerometers with sensitivity: 1.0 to 9.9 mV/(m/s2)

ACC (Acceleration): 1, 10, 100, 1000 m/s2

VEL (Velocity): 10, 100, 1000 mm/s

DISP (Displacement): 0.1, 1, 10, 100 mm

When accelerometer sensitivity is 0.1 to 0.99 mV/(m/s2), range increases by a factor of 10

When accelerometer sensitivity is 10 to 99 mV/(m/s2), range decreases by a factor of 1/10

— Indication characteristics Rion VM-82A

Acceleration: RMS, EQ PEAK

Velocity: RMS, EQ PEAK

Displacement: RMS, EQ PEAK, EQ p-p

EQ PEAK=RMS ×√2、EQ p-p=EQ PEAK × 2

— LCD panel (monochrome segment LCD)

Backlight: LED

Measurement value display: Display range 001 to 128; Mean value of 20 sampling values for each 100 ms is displayed, updated every 2 seconds

Bar graph display: Logarithmic scale, 1 to 100 % of full-scale

Indication characteristics: RMS, EQ PEAK, EQ p-p

Overload indication: “OVER” shown on display and screen color turns to red

Measurement mode indication: Acceleration, Velocity, Displacement

Memory address indication: 000 to 999 (1000 data)

Battery status indication: 4-segment display

Time indication: Year, month, day, hour, minute

Accelerometer sensitivity: 0.10 to 0.99, 1.0 to 9.9, 10 to 99 mV/(m/s2)

— Data memory

Maximum 1 000 data (000 to 999) can be stored manually

— Gain calibration

Accelerometer sensitivity selection establishes suitable gain

— Setting range Rion VM-82A

0.10 to 0.99, 1.0 to 9.9, 10 to 99 mV/(m/s2) (pC/(m/s2), when using VP-40/42)

— Output

          • AC output

Range full-scale: 1 V

Output impedance: Approx. 600 Ω

          • DC output

Range full-scale: 1 V

Output impedance: Approx. 600 Ω

          • Output voltage and display accuracy (electrical characteristics)

ACC (Acceleration): Range full-scale ±2 % (80 Hz)

VEL (Velocity): Range full-scale ±3 % (80 Hz)

DISP (Displacement): Range full-scale ±5 % (80 Hz)

         • Overall accuracy (in combination with PV-57エ)

ACC (Acceleration): Range full-scale ±5 % (80 Hz)

VEL (Velocity): Range full-scale ±8 % (80 Hz)

DISP (Displacement): Range full-scale ±10 % (80 Hz)

— Interfaces Rion VM-82A

USB: For data output and remote control of unit, data import to computer requires dedicated transfer software

— Ambient conditions for operation

Accelerometer: −20 °C to +70 °C, max. 90 % RH

Main unit: −10 °C to +50 °C, max. 90 % RH

— Power requirements

4 IEC R6 (size AA) batteries AC adapter (NC-98E, option)

— Current consumption: Approx. 65 mA
— Battery life (continuous use)

Alkaline batteries: Approx. 24 hours (room temperature, backlight OFF, outputs and communication function OFF)

Nickel-hydride batteries (eneloop XXR)*: Approx. 32 hours (room temperature, backlight OFF, outputs and communication function OFF)

— Dimensions / Weight

Approx. 171.5 (H) x 74 (W) x 25.5 (D) mm / Approx. 270 g (including batteries)

— Supplied accessories

Piezoelectric Accelerometer PV-57エ x 1, IEC LR6 (size AA) alkaline battery x 4, Curled accelerometer cable VP-51Kエ x 1, Magnet attachment VP-53S x 1

∗ Please use the dedicated charger to charged eneloop XXR batteries.

∗ eneloop XXR is a registered trademark of Panasonic group.

Features Rion VM-82A

– Operation panel with optimized button layout makes mode switching and setup easy and fast
– Wide range of measurement applications supported by selecting different accelerometers
– Backup function instantly reactivates previous settings at next power-on
– Convenient USB interface allows transfer of saved data to a computer
– Up to 24 hours of continuous operation on one set of alkaline button batteries. Environment-friendly nickel-hydride batteries are also supported.
– Compact dimensions and light weight: only 270 grams including batteries
Brand

,

Review RION VM-82A General-Purpose Vibration Meter

5 0% | 0 review
4 0% | 0 review
3 0% | 0 review
2 0% | 0 review
1 0% | 0 review
Review RION VM-82A General-Purpose Vibration Meter
Choose a image
0 character ( Minimum of 10)
    +

    There are no reviews yet.

    No comments yet

    RECENT PRODUCT

    hotline techmaster
    Hotline: 0908 173 345
    zalo techmaster Zalo: 0936 532 379 messenger techmaster Nhắn tin Facebook email techmaster Gửi Mail
    support techmaster