facebook

Ultra DI 20 Plus Liquid Optical Particle Counter

 

Features

  • Unmatched instrument to instrument data stability and matching with thermally stabilized measurements of particles across a 18 to 29°C operating environment
  • Instrument to instrument matching within 20% at concentrations of < 10 particles per ml
  • Compare UPW system performance across multiple fabs in any environmental conditions anywhere in the world
  • Fleet can be swapped without disruptions or questions about the data
  • Connect directly to PLC and SCADA systems with 4-20 mA and utilize existing network with Ethernet communication
Ultra DI 20 Plus Spec

Introduction PMS Ultra DI 20 Plus Liquid Optical Particle Counter

Sensitivity range: 20 – 100 nm

Designed specifically for ultrapure water systems, the Ultra DI 20 Plus Liquid Particle Counter from Particle Measuring Systems PMS quickly obtains statistically meaningful data and will count and size contaminants as small as 20 nm. Consistency helps you get more out of your data, even if the instrument moves, so you have more control over your process. The Ultra DI 20 Liquid Particle Counter has a low zero count, large sample volume, and high counting efficiency that provides benchmark particle detection. Improved sensitivity allows you to detect process variations more quickly with 20 nm polystyrene latex (PSL) and 9 nm gold (Au).

Particle Measuring Systems (PMS) is the world leader in sensitivity.

Ultra DI 20 Plus for ultrapure water

Features Ultra DI 20 Plus

  • Unmatched instrument to instrument data stability and matching with thermally stabilized measurements of particles across a 18 to 29°C operating environment
  • Instrument to instrument matching within 20% at concentrations of < 10 particles per ml
  • Compare UPW system performance across multiple fabs in any environmental conditions anywhere in the world
  • Fleet can be swapped without disruptions or questions about the data
  • Two counting modes
  • High resolution – for pure environments
  • High concentration – for filter challenge tests and lower purity environments
  • Connect directly to PLC and SCADA systems with 4-20 mA and utilize the existing network with Ethernet communication

Benefits Ultra DI 20 Plus

  • Advanced laser optics and sensors enable the detection of particles down to 20 nm (9 nm Au)
  • Quickly obtain meaningful statistical data to detect particle excursions and troubleshoot problems
  • Fast sample cleanup shortens the time to move from one sample point to another
  • Respond immediately to contamination with real-time particle measurement
  • Eliminates the additional purchase price and maintenance cost for an environmental enclosure

Applications

  • Quantifying particle concentration in state-of-the-art ultrapure water (UPW) systems
  • Filter efficiency measurements
  • Trending analysis at lower particle concentrations
  • Detecting bacterial growth in ultrapure water (UPW) systems
  • Episodic event tracking and alarming
  • Continuous system monitoring

Specifications

Channels 4
Channel sizes 20, 50, 70, 100 nm
Flow rate (ml/min) 75 ml/min ± 10%
Counting Efficiency 100% of sampling volume
Sample temperature 59 – 122 °F (15 – 50 °C)1
Maximum pressure 100 psi
Zero count ≤ 50 counts/L 2
Exterior surface 316L Stainless steel
Wetted surface materials Teflon®, Kel-F®, fused silica, Viton®, 96% Alumina Ceramic, and Simriz® 485
Dimensions (d, w, h) 20.0 x 16.8 x 14.5 in (50.8 x 42.7 x 36.8 cm)
Weight 71 lb (32.2 kg)
Communications Ethernet 4-20 mA (5 outputs: 4 particle channels, 1 instrument status) RS-232 (set up and diagnostics only, no data)
Status indicator Laser, power, and activity: one (1) tri-color LED
Calibration Materials used are traceable to National Institute of Standards and Technology (NIST)
Electrical Rating 100 – 240V, 50 – 60 Hz, 5.0 A Fuses 250 V, 5 x 20 mm, 5.0 A
Fuses 250 V, 5 x 20 mm, 5.0 A
Voltage fluctuation AC input voltage fluctuation shall not exceed ± 10%
Laser classification Class I complies with US21 CFR 1040.10 and EN60825-1. Internally an enclosed Class IV laser is used per EN60825-1. Operates with enclosure door open or closed
Environment Temperature: 64.4 – 84 °F (18 – 29 °C) ± 1 °C /hour 3

Humidity, Non-condensing

Maximum altitude: 6,562 ft (2,000 m)

Indoor use only

Pollution degree 2

Over-voltage Category II

Ordinary protection (Not protected against harmful ingress of moisture)

Class I environment (Electrical Earth ground from the mains power source to the product input is required for safety)

Storage and transport environment -40 – 158 °F (-40 – 70°C)

5 – 90 % Humidity

Learn more about the PMS Ultra DI® 20 Liquid Particle Counter version.

Products of Particle Measuring Systems.

Brand

Review Ultra DI 20 Plus Liquid Optical Particle Counter

5 0% | 0 review
4 0% | 0 review
3 0% | 0 review
2 0% | 0 review
1 0% | 0 review
Review Ultra DI 20 Plus Liquid Optical Particle Counter
Choose a image
0 character ( Minimum of 10)
    +

    There are no reviews yet.

    No comments yet

    RECENT PRODUCT

    hotline techmaster
    Hotline: 0936 532 379
    zalo techmaster Zalo: 0936 532 379 messenger techmaster Nhắn tin Facebook email techmaster Gửi Mail
    support techmaster