HSLIS M100e Chemical Particle Counter

 

Features

  • 4 channel chemical particle counter
  • 0.1, 0.2, 0.5 and 1.0 μm particle monitoring size range
  • Operates with an optimal flow rate of 300 ml/min. +/-10%
  • Samples 1% of the flow volume or 3 ml/min
  • Compatible with corrosive liquids, including hydrogen fluoride (HF)
  • eSeries compatible withFacility Net Software
  • Existing network can be used with ethernet communication to Facility Net Software
HSLIS-M100e Spec

Introduction HSLIS M100e Chemical Particle Counter

Sensitivity range: 0.1 – 1.0 μm

The High Sensitivity Liquid in Situ (HSLIS M100e) Chemical Particle Counter provides continuous, real-time particle monitoring of contamination levels in corrosive liquids, including hydrogen fluoride (HF) and is included in Particle Measuring System’ (PMS) complete line of liquid particle counters.

Particle Measuring Systems‘ HSLIS Liquid Particle Counter makes it easy to always understand the characteristics of your working fluid. The combination of high sensitivity and low cost makes it ideal when large numbers of sensors and/or long-term trending are required. Integration of the HSLIS Liquid Particle Counter into FacilityNet Software provides an economical way to monitor multiple flow systems and provide dedicated local display and control capabilities.

Chemical Particle Counter: HSLIS-M100e

Features HSLIS M100e Chemical Particle Counter

  • 4 channel chemical particle counter
  • 0.1, 0.2, 0.5 and 1.0 μm particle monitoring size range
  • Operates with an optimal flow rate of 300 ml/min. +/-10%
  • Samples 1% of the flow volume or 3 ml/min
  • Compatible with corrosive liquids, including hydrogen fluoride (HF)
  • eSeries compatible with Facility Net Software
  • Existing network can be used with ethernet communication to Facility Net Software

Benefits

  • Sensitivity range:  0.1 – 1.0 µm

Applications

  • Monitoring chemical delivery systems
  • Measuring filter efficiency
  • Episodic event tracking and alarming
  • Continuous chemical particle system monitoring

Specifications

Size range 0.1 – 1.0 µm
Channels 4
Channel sizes 0.1, 0.2, 0.5, 1.0 µm
Flow rate 300 ml/min ± 10%
Sample volume 3.0 ml/min
Counting efficiency 100% of sample volume
Maximum concentration 10,000 counts/ml
Sample temperature 194 °F (90 °C)
Zero count < 1000 counts/liter
Wetted surface materials Delrin®, Kel-F®, Teflon®, Viton®, fused silica
Dimensions (L x W x H) 21 x 10.75 x 7 in (53 x 27 x 18 cm)
Weight 28 lb (13 kg)
Power 85 – 250 VAC, 50 – 60 Hz
Calibration Materials used are traceable to US National Institute of Standards and Technology (NIST) and/or Japanese Institute of Standards (JIS).
Environment Temperature: 50 – 95 °F (10 – 35 °C); Humidity: non-condensing
Communications Ethernet, 4-20 mA (5 outputs; 4 particle channels, 1 instrument status)

RS-232 (set-up and diagnostics, no data)

Products of Particle Measuring Systems.

Brand

Review HSLIS M100e Chemical Particle Counter

5 0% | 0 review
4 0% | 0 review
3 0% | 0 review
2 0% | 0 review
1 0% | 0 review
Review HSLIS M100e Chemical Particle Counter
Choose a image
0 character ( Minimum of 10)
    +

    There are no reviews yet.

    No comments yet

    RECENT PRODUCT

    hotline techmaster
    Hotline: 0936 532 379
    zalo techmaster Zalo: 0936 532 379 messenger techmaster Nhắn tin Facebook email techmaster Gửi Mail
    support techmaster