Introduction HSLIS M100e Chemical Particle Counter
Sensitivity range: 0.1 – 1.0 μm
The High Sensitivity Liquid in Situ (HSLIS M100e) Chemical Particle Counter provides continuous, real-time particle monitoring of contamination levels in corrosive liquids, including hydrogen fluoride (HF) and is included in Particle Measuring System’ (PMS) complete line of liquid particle counters.
Particle Measuring Systems‘ HSLIS Liquid Particle Counter makes it easy to always understand the characteristics of your working fluid. The combination of high sensitivity and low cost makes it ideal when large numbers of sensors and/or long-term trending are required. Integration of the HSLIS Liquid Particle Counter into FacilityNet Software provides an economical way to monitor multiple flow systems and provide dedicated local display and control capabilities.
Features HSLIS M100e Chemical Particle Counter
- 4 channel chemical particle counter
- 0.1, 0.2, 0.5 and 1.0 μm particle monitoring size range
- Operates with an optimal flow rate of 300 ml/min. +/-10%
- Samples 1% of the flow volume or 3 ml/min
- Compatible with corrosive liquids, including hydrogen fluoride (HF)
- eSeries compatible with Facility Net Software
- Existing network can be used with ethernet communication to Facility Net Software
Benefits
- Sensitivity range: 0.1 – 1.0 µm
Applications
- Monitoring chemical delivery systems
- Measuring filter efficiency
- Episodic event tracking and alarming
- Continuous chemical particle system monitoring
Specifications
Size range | 0.1 – 1.0 µm |
Channels | 4 |
Channel sizes | 0.1, 0.2, 0.5, 1.0 µm |
Flow rate | 300 ml/min ± 10% |
Sample volume | 3.0 ml/min |
Counting efficiency | 100% of sample volume |
Maximum concentration | 10,000 counts/ml |
Sample temperature | 194 °F (90 °C) |
Zero count | < 1000 counts/liter |
Wetted surface materials | Delrin®, Kel-F®, Teflon®, Viton®, fused silica |
Dimensions (L x W x H) | 21 x 10.75 x 7 in (53 x 27 x 18 cm) |
Weight | 28 lb (13 kg) |
Power | 85 – 250 VAC, 50 – 60 Hz |
Calibration | Materials used are traceable to US National Institute of Standards and Technology (NIST) and/or Japanese Institute of Standards (JIS). |
Environment | Temperature: 50 – 95 °F (10 – 35 °C); Humidity: non-condensing |
Communications | Ethernet, 4-20 mA (5 outputs; 4 particle channels, 1 instrument status)
RS-232 (set-up and diagnostics, no data) |