facebook

HSLIS M50e Deionized Water Particle Counter

 

Features

  • 4 channel liquid particle counter
  • 0.05, 0.1, 0.15 and 0.2 μm size range
  • Operates with an optimal flow rate of 100 ml/min +/-10%
  • Samples 0.25% of the flow volume or 0.25 ml/min
  • eSeries compatible with Facility Net Software
  • Existing network can be utilized with ethernet communication to FacilityNet Software
HSLIS-M50e Spec

Introduction Deionized Water Particle Counter: HSLIS M50e

Sensitivity range: 0.05 – 0.2 µm

PMS Deionized Water Particle Counter HSLIS M50e provides continuous, real-time monitoring of contamination. The HSLIS Deionized Water Particle Counter makes it easy to always understand the characteristics of your working fluid. The combination of high sensitivity and low cost makes it ideal when large numbers of sensors and/or long-term trending are required. Integration of the HSLIS Liquid Particle Counter into Particle Measuring Systems’ Facility Net Software provides an economical way to monitor multiple flow systems and provide dedicated local display and control capabilities.

Deionized Water Particle Counter: HSLIS-M50e

Features Deionized Water Particle Counter: HSLIS M50e

  • 4 channel liquid particle counter
  • 0.05, 0.1, 0.15 and 0.2 μm size range
  • Operates with an optimal flow rate of 100 ml/min +/-10%
  • Samples 0.25% of the flow volume or 0.25 ml/min
  • eSeries compatible with Facility Net Software
  • Existing network can be utilized with ethernet communication to FacilityNet Software

Benefits

  • Monitor for long-term trends in fluid quality
  • Lower overall costs with the control and monitoring of multiple particle counters with Ethernet networks
  • Low initial capital investment and lower maintenance cost with proven reliability of the HSLIS design
  • Tabular and statistical process control (SPC) charts for detailed statistical information
  • Alphanumeric paging and e-mailing functions to alert operator of data excursions
  • Sensor status displays four-level alarm for monitoring out-of-spec conditions
  • Simplified data interpretation with real-time and retrieved data time plots

Applications

  • Quantifying particle concentrations in DI water systems
  • Measuring filter efficiency
  • Detecting bacterial growth in DI water systems
  • Episodic event tracking and alarming
  • Continuous system monitoring

Specifications

Size range 0.05 – 0.2 µm
Channels 4
Channel sizes 0.05, 0.1, 0.15, 0.2 µm
Flow rate 100 ml/min ± 10%
Sample volume 0.25 ml/min
Counting efficiency 100% of sample volume
Maximum concentration 10,000 counts/ml
Sample temperature 194 °F (90 °C)
Zero count < 1000 counts/liter
Wetted surface materials Delrin®, Kel-F®, Teflon®, Viton®, fused silica
Dimensions (L x W x H) 21 x 10.75 x 7 in (53 x 27 x 18 cm)
Weight 28 lb (13 kg)
Power 85 – 250 VAC, 50 – 60 Hz
Calibration Materials used are traceable to US National Institute of Standards and Technology (NIST) and/or Japanese Institute of Standards (JIS).
Environment Temperature: 50 – 95 °F (10 – 35 °C); Humidity: non-condensing
Communications Ethernet, 4-20 mA (5 outputs; 4 particle channels, 1 instrument status)

RS-232 (set-up and diagnostics, no data)

Products of Particle Measuring Systems.

Brand

Review HSLIS M50e Deionized Water Particle Counter

5 0% | 0 review
4 0% | 0 review
3 0% | 0 review
2 0% | 0 review
1 0% | 0 review
Review HSLIS M50e Deionized Water Particle Counter
Choose a image
0 character ( Minimum of 10)
    +

    There are no reviews yet.

    No comments yet

    RECENT PRODUCT

    hotline techmaster
    Hotline: 0936 532 379
    zalo techmaster Zalo: 0936 532 379 messenger techmaster Nhắn tin Facebook
    support techmaster