Introduction to the Yokogawa CA71 Multifunction Calibrator
The Yokogawa CA71 Multifunction Calibrator is a compact, lightweight, and versatile device capable of simultaneously sourcing and measuring voltage, current, resistance, thermocouples (TC), resistance temperature detectors (RTD), frequency, and pulse signals with high accuracy. It features easy-to-use rotary switches, making it ready for immediate use by simply opening the carrying case and connecting the cables.
Features of the Yokogawa CA71 Multifunction Calibrator
- Power Source: The CA71 provides 4-20 mA steps and 24V DC power for transmitters.
- Output Divide Function: Simplifies percentage output calculations.
- Auto Step Function: Changes output values incrementally based on the divide output function settings.
- Online Communication (CA71): Features an optically isolated RS-232C interface.
- Scan Function: Increases or decreases output linearly with selectable rise/fall times of 16 or 32 seconds.
- Memory Function: Stores and recalls up to 50 sets of source and measurement values from the internal memory.
- Temperature Monitoring Function
Specifications Summary – Model CA71
Functions | Details |
---|---|
Auto Step Function | 2.5 sec/step or 5 sec/step |
Scan Function | 16 sec or 32 sec |
Memory Function | Stores 50 value sets |
Maximum Measurement Input | Voltage: 300 V AC, Current: 120 mA DC |
Current Input Terminal Protection | Fuse: 125 mA/250 V |
Earth Voltage of Measurement Unit | Maximum 300 V AC |
Display Update Rate | Approximately once per second |
Serial Interface | RS232 connection (sold separately) |
Display | Segment LCD (76 mm × 48 mm) |
Backlight | LED backlight, auto-off after 1 minute |
Power Supply | 4 AA alkaline batteries or AC adapter (sold separately) |
Battery Life | 12 to 40 hours (depending on usage conditions) |
Auto Power Off Function | Approximately 10 minutes (can be disabled) |
Applicable Standards | EN55011: 1998, EN61326-1: 1997 + A1: 1998, IEC61010-1, IEC61010-2-31 |
Insulation Resistance | 500 V DC, 50 MΩ or higher |
Withstand Voltage | 3.7 kVAC, for 1 minute |
Operating Temperature and Humidity | 0 – 50°C, 20 – 80% RH (non-condensing) |
Storage Temperature and Humidity | -20 – 50°C, 90% RH or lower (non-condensing) |
Dimensions | 190 × 120 × 55 mm |
Weight | 730 g (including batteries) |
Reviews
There are no reviews yet.