Fluke 730G Digital Pressure Calibrator
Product Overview
A compact, smart pressure solution for bench or field
The Fluke 730G Smart Digital Pressure Calibrator combines the performance of a precision test gauge with the power of a loop calibrator and also includes internal HART communication capabilities. When used with a pressure test pump the Fluke 730G becomes a high-efficiency, highly accurate calibrator that can test pressure gauges, pressure switches and pressure transmitters making it ideal for instrument shops, gauge labs and pressure laboratories.
- Built-in HART Communication: The integrated HART communication functionality is easy-to-use and allows you to adjust HART directly, without requiring additional tools for adjustment.
- Precision Pressure and Current Measurements: The 730G series provides high-precision 0.02% pressure measurement accuracy and 0.01% current measurement accuracy to meet the requirements for establishing a pressure laboratory and achieving accurate pressure calibration results.
- Intuitive operation and ease of use: A high-end LCD display and dual-jack design bring a more convenient and efficient user experience.
- Paperless testing and reporting: Optional 730G/PCSW management software enables paperless testing giving you the ability to automatically generate test reports and document results.
Fluke 730G/PCSW Software
- Paperless management of pressure calibrations: Record calibration data, automate calibration tests and generate calibration reports.
- Asset management: Manage tested equipment and pressure standards with ease.
- Supported pressure devices include: analog pressure gauges, digital pressure gauge, pressure transmitter, pressure sensors and pressure switches.
Increase your pressure testing capabilities using test pump kits
Fluke-700PTPK Pneumatic test pump kit:
- Combine with any Fluke-730G Series Gauge, 1000 psi (69 bar) or less
- Generate pressure up to 600 psi (40 bar) with the 700PTP-1 test pump
- Connect the 730G Series gauge directly to the included PTP-1 hand pump
- Includes hoses and adapters for connecting to the pressure device to be tested
- Hard case protects pump and gauge and allows gauge to remain connected to the pump
Fluke-700HTPK Hydraulic Test Pump Kits
- Combine with any Fluke-730G Series Gauge, 1000 psi (69 bar) or greater
- Generate pressure up to 10000 psi (690 bar) with the 700HTP-2 test pump
- Connect the 730G Series gauge directly to the HTP-2 hand pump
- Includes 700HTH-1 hoses and adapters for connecting to the pressure device to be tested.
- Hard case protects pump and gauge and allows gauge to remain connected to the pump
Specifications: Fluke 730G Digital Pressure Calibrator
Technical specifications | ||||
Pressure specifications: 0.02% of full scale (730RG), 0.05% of full scale (730G) | ||||
Current specifications: 0.01% of reading | ||||
Model | Range | Resolution | Accuracy1 | Burst pressure |
730G01 | -10 to +10 inH2O -20 to 20 mbar -2.49 to 2.49 kPa |
0.001 inH2O 0.001 mbar 0.0001 kPa |
Positive pressure, ±0.2% of range2 Vacuum, ±0.2% of range2 |
3 psi, 210 mbar 20.6 kPa |
730G02 | -1 to +1 psi -70 to 70 mbar -6.89 to 6.89 kPa |
0.0001 psi 0.001 mbar 0.0001 kPa |
5 psi, 350 mbar 34.4 kPa |
|
730G04 | -14 to 15 psi -0.97 to 1 bar -97 to 103 kPa |
0.001 psi 0.0001 bar 0.01 kPa |
Positive pressure, ±0.05% of range Vacuum, ±0.1% of range |
60 psi 4 bar 413.6 kPa |
730G05 | -14 to 30 psi -0.97 to 2 bar -97 to 206 kPa |
0.001 psi 0.0001 bar 0.01 kPa |
120 psi 8 bar 827.3 kPa |
|
730G06 | -12 to 100 psi -0.83 to 6.9 bar -83 to 690 kPa |
0.01 psi, 0.0001 bar | 400 psi 26 bar 2.757 MPa |
|
730G27 | -12 to 300 psi -0.83 to 20 bar -83 to 2 MPa |
0.01 psi 0.001 bar 0.1 kPa |
1200 psi 80 bar 8.273 MPa |
|
730G07 | -12 to 500 psi -0.83 to 34 bar -83 to 3.4 MPa |
0.01 psi 0.001 bar 0.1 kPa |
2000 psi 138 bar 13.789 MPa |
|
730G08 | -14 to 1000 psi, -0.97 to 69 bar -97 to 6900 kPa |
0.01 psi 0.001 bar 0.1 kPa |
4000 psi 266 bar 27.579 MPa |
|
730G10 | -14 to 2000 psi, -0.97 to 140 bar -97 to 14 MPa |
0.1 psi 0.01 bar 1 kPa |
8000 psi 550 bar 55.158 MPa |
|
730G29 | -14 to 3000 psi, -0.97 to 200 bar -97 to 20 MPa |
0.1 psi 0.01 bar 1 kPa |
10000 psi 690 bar 68.947 MPa |
|
730G30 | -14 to 5000 psi, -0.97 to 340 bar -97 to 34 MPa |
0.1 psi 0.01 bar 1 kPa |
10000 psi 690 bar 68.947 MPa |
|
730G31 | -14 to 10000 psi, -0.97 to 690 bar -97 to 69 MPa |
1 psi 0.01 bar 1 kPa |
15000 psi 1035 bar 103.421 MPa |
|
730GA4 | 0 to 15 psia, 0 to 1 bar 0 to 100 kPa absolute pressure |
0.001 psi 0.0001 bar 0.01 kPa |
±0.05% of range4 | 60 psi 4 bar 413.6 kPa |
730GA5 | 0 to 30 psia 0 to 2 bar 0 to 200 kPa absolute pressure |
0.001 psi 0.0001 bar 0.01 kPa |
120 psi 8 bar 827.3 kPa |
|
730GA6 | 0 to 100 psia, 0 to 6.9 bar 0 to 690 kPa absolute pressure |
0.01 psi 0.0001 bar 0.01 kPa |
400 psi 27 bar 2.757 MPa |
|
730GA27 | 0 to 300 psia 0 to 20 bar 0 to 2 MPa absolute |
0.01 psi 0.001 bar .01 kPa |
1200 psi 80 bar 8.273 MPa |
|
730RG04 | -14 to 15 psi, -0.97 to 1 bar -97 to 100 kPa |
0.001 psi 0.0001 bar 0.0001 kPa |
Positive pressure, ±0.02 % of range3 Vacuum, ±0.05 % of range3 |
45 psi 3 bar 310.2 kPa |
730RG05 | -14 to 30 psi, -0.97 to 2 bar -97 to 200 kPa |
0.001 psi 0.0001 bar 0.01 kPa |
90 psi 6 bar 620.5 kPa |
|
730RG06 | -12 to 100 psi, -0.83 to 6.9 bar -83 to 690 kPa |
0.01 psi 0.0001 bar 0.01 kPa |
1000 psi 69 bar 6.894 MPa |
|
730RG27 | –12 psi to 300 psi -0.83 to 20 bar –80 kPa to 2 MPa |
0.01 psi 0.001 bar 0.1 kPa |
2000 psi 138 bar 14 MPa |
|
730RG07 | –12 psi to 500 psi -0.83 to 34 bar –80 kPa to 3.5 MPa |
0.01 psi 0.001 bar 0.1 kPa |
2000 psi 138 bar 13.789 MPa |
|
730RG08 | -14 to 1000 psi, -0.97 to 69 bar -97 to 6900 kPa |
0.1 psi 0.001 bar 0.1 kPa |
10000 psi 690 bar 68.947 MPa |
|
730RG29 | -14 to 3000 psi -0.97 to 200 bar -97 to 20 MPa |
0.1 psi 0.01 bar 1 kPa |
10000 psi 690 bar 68.947 MPa |
|
730RG30 | -14 to 5000 psi, -0.97 to 340 bar -97 to 34 MPa |
0.1 psi 0.01 bar 1 kPa |
10000 psi 690 bar 68.947 MPa |
|
730RG31 | -14 to 10000 psi, -0.97 to 690 bar -97 to 69 MPa |
0.1 psi 0.01 bar 1 kPa |
15000 psi 1035 bar 103.421 MPa |
|
1 Accuracy specifications above have considered the effects of linearity, repeatability, resolution, reference accuracy, 1-year stability, and temperature effect. 2 For 730G01/730G02, accuracy is ±0.3% of full scale @ 0 °C to 15 °C and 35 °C to 50 °C. 3 For all 730RG model, add 0.003% FS/°C @ 0 °C to 18 °C and 28 °C to 50 °C。 4 For 730GA4, add 0.006% FS/°C @ -10 °C to 15 °C and 35 °C to 55 °C; for all other 730GA model, add 0.003% FS/°C @ -10 °C to 15 °C and 35 °C to 55 °C. |
||||
Media Compatibility 730G01, 730G02, 730G04, 730G05, 730RG04, 730RG05: any clean dry non-corrosive gas Above 100 psi: any liquids or gases compatible with 316 stainless steel |
Electrical specifications (All specifications are valid to 110% of range) | |
Current | 0 mA to 24 mA |
Voltage | 0 V DC to 30 V DC |
Current range | 1 μA |
Voltage range | 1 mV |
Accuracy | 0.01% ± 2 LSD all ranges @ 23°C ±5 °C |
Temperature compensation | From 20% FS/°C @ -10 °C to +18 °C and 28 °C to 50 °C |
Loop Compliance Voltage | 24 V DC @ 20 mA |
Temperature measurement accuracy inside the product | ±0.5 ℃ |
Drive Capability | 1200 Ω (without HART resistor), 950 Ω (with internal HART resistor) |
Review Fluke 730G Digital Pressure Calibrator
There are no reviews yet.