Introduction Fluke 726 Precision Multifunction Calibrators
Process calibration designed specifically for process industries
The Fluke 726 is a precise and powerful, yet easy-to-use, field calibrator. It features broad workload coverage, calibration power, and unsurpassed accuracy needed by process professionals. It measures and sources almost all process parameters to calibrate almost anything. Use it to test sensors and valves, and to test and calibrate transmitters. The 726 interprets results without the help of a calculator and stores measurement data for later analysis.
Measurement capabilities
This precision multifunction calibrator measures volts, mA, pressure, RTDs, thermocouples, frequency, and resistance to test sensors and transmitters. The 726 also sources and simulates volts, mA, thermocouples, RTDs, frequency to calibrate transmitters. Measures and/or sources pressure using when paired with any of the 50 Fluke 750P series Pressure Modules to test and calibrate pressure transmitters and other devices. The 726 can also:
- Source mA with simultaneous pressure measurement to conduct valve and I/P test
- Conduct enhanced flowmeter testing with frequency totalizer and frequency pulse train source mode
Designed for ease of use
The 726 has built-in features that make it easier to use in nearly any measurement situation. From its ability to store frequently used test setups, to programming custom RTD probe calibration constants for enhanced temperature measurement, or the integrated voltage input protection design that improves reliability, the 726 is built to perform. The use of two separate channels even allows you to measure, source, or view process signals simultaneously. The backlit display provides better visibility in low-light conditions, and the standard 3-year warranty provides extra confidence that the 726 will be there when you need it to be.
Features
- Precision multifunction, temp, frequency, voltage, and mA calibrator
- Calibrates thermocouples and RTD temperature instruments
- Stores up to eight calibration results in memory for analysis
- Built-in 24 V Loop Power Supply
- Capture the set, reset, and deadband of a pressure switch with the integrated switch test
- Calculate the transmitter error % to interpret calibration results without the use of a calculator
- Perform fast linearity tests with the auto-step and auto-ramp features
- Power a transmitter during test using 24 V loop supply and simultaneous mA measurement
Specifications
Measurement Accuracy | ||
Voltage DC | 30.000 V | 0.01% + 2 mV (upper display) |
20.000 V | 0.01% + 2 mV (lower display) | |
90.00 mV | 0.01% + 10 μV | |
-10.00 mV to 75.00 mV | 0.025% + 1 count (via TC connector) | |
Current DC | 24.000 mA | 0.01%+ 2 mA |
Resistance | 0.0 to 4000.0 Ω | ± 0.015% |
Frequency | 2.0 to 1000.0 CPM | 0.05% + 0.1 cpm |
1.0 to 1000.0 Hz | 0.05% + 0.1 Hz | |
1.00 to 15.00 kHz | 0.05% + 0.01 kHz | |
Sensitivity | 1 V peak-to-peak-minimum | |
Pressure | Accuracy from 0.025% of range using any of 29 pressure modules (for detailed specifications refer to pressure modules in options and accessories). Modules available for differential, gauge, vacuum, absolute, dual and high pressure. | |
Source Accuracy | ||
Voltage DC | 100.00 mV | 0.01% + 10 μV |
20.000 V | 0.01% + 2 mV | |
-10.00 mV to 75.00 mV | 0.025% + 1 count (via TC connector) | |
Current DC | 24.000 mA (Source) | 0.01% + 2 μA |
24.000 mA (Simulate) | 0.01% + 2 μA | |
Resistance | Range, accuracy (percent of reading + floor) and excitation current | 5 to 400 Ω, 0.015% + 0.1 Ω, 0.1 to 0.5 mA |
5 to 400 Ω, 0.015% + 0.05 Ω, 0.5 to 3 mA | ||
400 to 1500 Ω, 0.015% + 0.5 Ω, 0.05 to 0.8 mA | ||
1500 to 4000 Ω, 0.015% + 0.5 Ω, 0.05 to 0.4 mA | ||
Frequency | 2.0 to 1000.0 CPM | 0.05% |
1.0 to 1000.0 Hz | 0.05% | |
1.0 to 10.0 kHz | 0.25% | |
10.0 to 15.0 kHz | 0.5% | |
Waveform | 5 V p-p squarewave, -0.1 V offset | |
RTDs and Thermocouples | ||
Measure and source accuracy | Cu10 | 1.8°C |
NI-120 | 0.15°C | |
PT-100 (385) | 0.45°C | |
PT-100 (392) | 0.35°C | |
PT-100 (391-JIS) | 0.35°C | |
PT-200 (385) | 0.95°C | |
PT-500 (385) | 0.55°C | |
PT-1000 (385) | 0.35°C | |
RTD Resolution | 0.01°C (0.01°F) | |
Cu10.01 C | 0.1°C (0.1°F) | |
J | 0.4°C | |
K | 0.5°C | |
T | 0.4°C | |
E | 0.4°C | |
R | 1.4°C | |
S | 1.4°C | |
B | 1.4°C | |
L | 0.45°C | |
U | 0.45°C | |
N | 0.6°C | |
XK | 0.4°C | |
BP | 1.1°C | |
TC Resolution | J, K, T, E, L, N, U: 0.1°C, 0.1°F B, R, S: 1°C, 1°F | |
Specifications | ||
Ramp functions | Source functions | Voltage, current, resistance, frequency, temperature |
Ramps | Slow ramp, Fast ramp, 25% step-ramp | |
Loop power function | Voltage | 24 V |
Accuracy | 10% | |
Maximum current | 22 mA, short circuit protected | |
Step functions | Source functions | Voltage, current, resistance, frequency, temperature |
Steps | 25% of range, 100% of range | |
Environmental Specifications | ||
Operating temperature | -10°C to +55°C | |
Storage temperature | -20°C to +71°C | |
Operating altitude | 3000 m | |
Safety Specifications | ||
Agency approvals | EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992 | |
Mechanical and General Specifications | ||
Size | 200 x 96 x 47 mm (7.9 x 3.8 x 1.9 in) | |
Weight | 650 g (23 oz) | |
Batteries | 4 AA alkaline batteries | |
Battery replacement | Separate battery compartment, accessible without breaking calibration seal | |
Side port connections | Pressure module connector, also used for remote real-time programming |
Review Fluke 726 Precision Multifunction Calibrators
There are no reviews yet.