SPECIFICATION HIOKI DT4256 DIGITAL MULTIMETER
DC Voltage range | 600.0 mV to 1000 V, 5 ranges Basic accuracy: ±0.3 % rdg. ±3 dgt. |
---|---|
AC Voltage range | 6.000 V to 1000 V, 4 ranges, Frequency characteristics: 40 Hz to 1 kHz Basic accuracy 40 – 500 Hz : ±0.9 % rdg. ±3 dgt. (True RMS, crest factor 3) |
AUTO AC/DCV | Yes |
Resistance range | 600.0 Ω to 60.00 MΩ, 6 ranges, Basic accuracy: ±0.7 % rdg. ±3 dgt. |
DC Current range | 60.00 mA to 10.00 A, 4 ranges, Basic accuracy: ±0.9 % rdg. ±3 dgt. |
AC Current range |
600.0 mA to 10.00 A, 3 ranges, Basic accuracy 40 – 500 Hz : ±1.4 % rdg. ±3 dgt. (True RMS, crest factor 3, 40 Hz to 1 kHz) |
AC Current range (use with Clamp on probes) | 10.00 A to 1000 A, 7 ranges, Add the Clamp on probe accuracy to basic accuracy 40 – 1 kHz : ±0.9 % rdg. ±3 dgt. (True RMS, crest factor 3) |
Voltage detection (50/60 Hz) | Hi: AC40 V to 600 V, Lo: AC80 V to 600 V |
Capacitance range | 1.000 μF to 10.00 mF, 5 ranges, Basic accuracy: ±1.9 % rdg. ±5 dgt. |
Frequency range | 99.99 Hz (5 Hz or more) to 99.99 kHz, 4 ranges (limited by the minimum detectable voltage and current), Basic accuracy: ±0.1 % rdg. ±1 dgt. |
Continuity check | Continuity threshold [ON]: 25 Ω or less (Indicate buzzer sound, red LED), [OFF]: 245 Ω or more, Response time: 0.5 ms or more |
Diode test |
Open terminal voltage: 5.0 V or less, Testing current 0.5 mA or less, Threshold of forward voltage: 0.15 V to 1.5 V |
Other functions | Filter function, display value hold, auto hold, MAX/MIN/Average value display, relative value display, auto-power save, USB communication (option) |
Display | Main and Sub displays: 4-digits LCD, max. 6000 digits, bar graph |
Display refresh rates | 5 times/s (Capacitance measurement: 0.05 to 5 times/s, depending on measured value, Frequency: 1 to 2 time/s, Temperature: 1 time/s ) |
Power supply | LR03 alkaline batteries ×4, Continuous use: 130 hours (backlight OFF) |
Dimensions and mass | 84 mm (3.31 in)W × 174 mm (6.85 in)H× 52 mm (2.05 in)D, 390 g (13.8 oz) (including batteries and holster) |
Accessories | Test lead L9207-10 ×1, Holster ×1, Instruction manual ×1, LR03 alkaline battery ×4 |
Introduction Hioki DT4256 Digital Multimeter
The Hioki DT4200 Digital Multimeter Series consist of a complete lineup of 9 models ranging from professional to industrial to pocket meters, delivering a super fast response rate and safety features that take electrical testing to a higher level. The Hioki DT4256 is a digital multimeter designed to deliver maximum functionality for use in a wide range of settings. CAT IV 600 V, CAT III 1000 V.
Key Features Hioki DT4256 Digital Multimeter
• ±0.3% DC V basic accuracy, wide 40 Hz to 1 kHz AC V frequency characteristics
• Measure up to 10A with direct input
• Dual display lets you check voltage and frequency simultaneously
• Low-pass filter cuts high harmonics (when measuring inverter fundamental waveforms)
• USB communications function supports PC measurements (optional)
• Broad -25 (-13°F) to 65°C (149°F) operating temperature range (DT4256)
Hioki Brand
Established in 1935, HIOKI E.E. CORPORATION (TSE: 6866) has grown to become a world leader in providing consistent delivery of test and measuring instruments through advanced design, manufacturing, and sales and services. Hioki test tools are trusted for their safety, reliability and quality by electrical professionals for use in rigorous environments around the world. Hioki products apply the latest and most advanced technologies to ensure the highest accuracy and safety.
Review HIOKI DT4256 (True RMS) DIGITAL MULTIMETER