facebook

Calibration of torque measuring Instruments by the Snap-on TTC2000 system

Calibration of torque measuring Instruments by the Snap-on TTC2000 system

In the current market, permissible error for torque calibration systems is commonly 1% of measured value. However, The Snap-On TTC2000 system has a low permissible error of 0.25% of the measured value.

Explore torque measuring tool calibration with Snap-On TTC2000 system with Techmaster (Calibration of torque measuring Instruments by the Snap-on TTC2000 system).

1. The calibration steps

  • Calibration of torque measuring Instruments by the Snap-on TTC2000 system

1.1 Calibration equipment

  • Standard equipment

– Snap-On TTC2000 Electronic Torque Calibration System

– The transducers corresponding to the measurement ranges are as follows: (0.45-5.65) N.m, (27.1-339) N.m, (81.3-813) N.m, (271-2710) N.m

  • The Measuring instruments to be verified

– The torque measuring tools within the calibration range of the SNAP-ON TTC2000 calibration system.

1.2 Calibration environmental conditions 

+ Calibration conditions for Torque using the Snap-On TTC2000 system.

– Temperature: (23 ± 5) ºC, with a stability of ± 2 ºC during the calibration process.

– Humidity: ≤ 80% RH, with a stability of ± 5% RH during the calibration process.

1.3 Perform calibration

– Starting up the electronic torque calibration system

– Select a standard sensor with a suitable measurement range for the measuring instrument being checked.

1.3.1 Perform torque measurement verification within the ranges of (0.45-5.65) newton-met and (27.1-339) newton-met using the corresponding measuring instruments

– Attach instrument to calibration system per manufacturer’s instructions.

– Adjust the measurement unit of the calibration system to match the measurement unit of the measuring instrument. 

– Verify at 5 evenly spaced points as required by the customer, with 5 repetitions per point.

– Perform verification in both clockwise (CW) and counterclockwise (CCW) directions. If the measuring instrument is single-directional, verify only in that specific direction.

– Print the results after each measurement.

– Compare the average measured values from the calibration system with the set or displayed values of the torque measuring instrument.

Perform calibration

1.3.2 Verify the measuring instrument within the range of (81.3-813) newton-met

– Replace the transducer with the corresponding measurement range mentioned above.

– Perform a similar verification following the same steps as mentioned above.

1.3.3 Verify the measuring instrument within the range of (81.3-813) newton-met

– Replace the transducer with the corresponding measurement range.

– Repeat the same verification process as mentioned above.

Leave a Reply

Your email address will not be published.

hotline techmaster
Hotline: 0936 532 379
zalo techmaster Zalo: 0936 532 379 messenger techmaster Nhắn tin Facebook
support techmaster