LCR Meter is a specialized device for testing the inductance (L), capacitance (C) and resistance (R) of electronic components. This equipment is applied at laboratories, factories, warranty offices to research, develop, testand repair electronic components. The LCR Meter series possesses the ability to measure a variety of parameters: the inductor quality coefficient (Q) and the coefficient (D) is similar to Q but for resistance. With the improvement in technology, LCR Meter is integrated with other measuring devices such as oscilloscopes, pulse generators, … Provides many parameters for users to accurately assess the device status. The testing efficiency, maintenance and repair work will achieve higher efficiency.
2. Scope of procedure
Unit Under Test Characteristics | Performance Range | Performance method |
LCR Meter Handhelds LCR Meter | Up to 100 MΩ Up to 100 μF Up to 10 H (20 Hz ~ 1 MHz) | Comparison to by LCR Standard |
✔ Digital Multimeter
✔ LCR Calibrator
✔ Standard Inductor
✔ Standard Capacitor
✔ Decade Resistor
✔ Connecting Leads/Wires
✔ LCR meter test leads / 4 BNC Connector
Temperature: (18 ÷ 28)°C
Relative Humidity: (40 ÷ 60)%RH
– Connect UUT as show as above.
– Calibration points will be perform at 100 Hz, 120 Hz, 1kHz, 10kHz, 100kHz with 1V. Other test point follow customer’s requirements.
– Set Digital Multimeter function switch to measure frequency.
– Adjust UUT Frequency for the desired output value.
– Record the Digital Multimeter indicator.
– Repeat at least 5 times
– Repeat for remain test points.
– Disconnect test leads.
Frequency | UUT Setting |
1kHz | 1 Vrms |
2 Vrms | |
3 Vrms | |
4 Vrms | |
5 Vrms | |
DC | 0.5 V |
1 V | |
1.5 V | |
2 V | |
3 V |
– Connect UUT.
– Set UUT function measure Capacitance.
– Set UUT with signal level: 1Vrms , Trigger : Int, Speed : slow , Cable Length 1m.
Frequency | Capacitor Value |
20 Hz ~ 1 MHz | 10 pF |
100 pF | |
1 nF | |
10 nF | |
100 nF | |
1 μF | |
10 μF | |
100 μF |
– Connect UUT as show Fig.2 for 4 BNC connector.
– Connect UUT as show Fig.3 for LCR meter test leads.
– Set UUT function measure Resistance.
– Set UUT with signal level: 1Vrms , Trigger : Int, Speed : slow .
– Set UUT range dependent on the appropriate resistance in table below. Other test point follow customer’srequirements.
Frequency | Resistor Value |
20 Hz ~ 1 MHz | 0.1 mΩ |
1 mΩ | |
10 mΩ | |
100 mΩ | |
1 Ω | |
10 Ω | |
100 Ω | |
1 kΩ | |
10 kΩ | |
100 kΩ | |
1 MΩ | |
10 MΩ | |
100 MΩ |
– Connect UUT as show Fig.2 for 4 BNC connector.
– Connect UUT as show Fig.4 for LCR meter test leads.
– Set UUT function measure Inductance.
– Set UUT with signal level: 1Vrms , Trigger : Int, Speed : slow .
– Set UUT range dependent on the appropriate inductor in table below. Other test point follow customer’s requirements.
Frequency | Inductor Value |
20 Hz ~ 1 MHz | 10 μH |
100 μH | |
1 mH | |
10 mH | |
100 mH | |
1 H | |
10 H |
– Record the UUT value.
– Repeat at least 5 times.
– Repeat the above steps with other frequency points.
– Disconnect the inductancefrom UUT test leads.
Calibration now completed, turn off, disconnect and secure all equipment.
*Note : Other function of equipment reference to the manufacturer’s instructions or customer’s requirement.
– Customer required tolerances are to be used for all calibrations unless otherwise noted.
– If the customer does not supply instrument tolerances then manufacturing specifications or other approved sources are to be used as the default tolerance.
– If not, non-judgement results will be applied
New Techmaster Headquarters: Strategic Vision at a Prime Location In the dynamic context of Vietnam’s…
Techmaster and Keysight Technologies Announce strategic partnership in Vietnam Techmaster is proud to announce its…
When working in toxic or low-oxygen environments, safety begins with proper respiratory protection. The 3M…
Vertical accuracy is a key factor that determines the measurement quality of an oscilloscope. It…
In the latest update of the Scope of Accreditation to ISO/IEC 17025:2017 and ANSI/NCSL Z540-1-1994…
ISO 17025 certification ensures that laboratories meet international standards for testing and calibration competence. It…