1. Scope of procedure

Unit Under Test Characteristics

Performance Range

Performance method

Temperature

(-10 ÷ 500)°C

Comparison to Blackbody.

2. Equipment, standard requirements

–  Blackbody: FLUKE, OMEGA or equivalent, accuracy better than 3 times UUT.

– Measuring tape.

3. Environment conditions

Temperature: (20 ÷ 28)°C

Relative Humidity: (30 ÷ 70)%RH

4. Preparation

– Visually inspect the Unit Under Test (UUT) for obvious damage that may affect accuracy and/or operation and note any observations.
– Connect UUT to suitable power and turn on.
– Turn on Blackbody, allow warm-up at least 10 minutes or follow manufacturer’s instructions.
– Technician must be familiar with calibration techniques, content of this calibration procedure and the operator manual of the UUT being calibrated, and the generator being used.

5. Perform

–  Almost infrared thermometer has default emissivity 0.95. Some IR device such as thermal imager, thermal camera can adjust other emissivity. In this case, emissivity shall be set at 0.95 also.
– Follow user guide of Blackbody, set emissivity equal to 0.95, same with UUT.
– Base on UUT Distance to Spot ratio (D:S), use measuring tape to set the distance between Blackbody emitting surface and the UUT, that required for calibration process.
– Perform at least 5 points evenly distributed across the UUT range. Other test points may follow customer’s requirements.
– Begin at the lowest calibration point, example -10oC first, then increase for higher points.
– Set the Blackbody for the first calibration point and allow sufficient time to stabilize.
– Waiting for stable reading on UUT and the Blackbody, record repeat values.
– Repeat for remain test points.
– Calibration now completed, set and allow temperature of Blackbody near to normal room temperature.
– Turn off Blackbody and UUT, take out and secure all equipment.

6. Tolerance

– Tolerance given by the customer will be applied for all measurement, unless otherwise provisions.
– If the customer does not required, the specifications of manufacturer or other source of reference will be used.
– If not, the tolerance ±2oC or ±2%reading of UUT will be applied.
Nguyen Thuy

Share
Published by
Nguyen Thuy

Recent Posts

TECHMASTER’s New Headquarters: A Strategic Vision at a Prime Location – 175 Vo Nguyen Giap

New Techmaster Headquarters: Strategic Vision at a Prime Location In the dynamic context of Vietnam’s…

2 months ago

Techmaster is Keysight Technologies authorized distributor in Vietnam

Techmaster and Keysight Technologies Announce strategic partnership in Vietnam Techmaster is proud to announce its…

6 months ago

3M Scott SCBA: Breathe Safe in Any Situation

When working in toxic or low-oxygen environments, safety begins with proper respiratory protection. The 3M…

6 months ago

Vertical accuracy in oscilloscope

Vertical accuracy is a key factor that determines the measurement quality of an oscilloscope. It…

10 months ago

ISO/IEC 17025:2017 Scope Updates v015

In the latest update of the Scope of Accreditation to ISO/IEC 17025:2017 and ANSI/NCSL Z540-1-1994…

1 year ago

Vietnam Businesses: ISO 17025

ISO 17025 certification ensures that laboratories meet international standards for testing and calibration competence. It…

1 year ago