1. Scope of procedure

Unit Under Test Characteristics

Performance Range

Performance method

Temperature

(-10 ÷ 500)°C

Comparison to Blackbody.

2. Equipment, standard requirements

–  Blackbody: FLUKE, OMEGA or equivalent, accuracy better than 3 times UUT.

– Measuring tape.

3. Environment conditions

Temperature: (20 ÷ 28)°C

Relative Humidity: (30 ÷ 70)%RH

4. Preparation

– Visually inspect the Unit Under Test (UUT) for obvious damage that may affect accuracy and/or operation and note any observations.
– Connect UUT to suitable power and turn on.
– Turn on Blackbody, allow warm-up at least 10 minutes or follow manufacturer’s instructions.
– Technician must be familiar with calibration techniques, content of this calibration procedure and the operator manual of the UUT being calibrated, and the generator being used.

5. Perform

–  Almost infrared thermometer has default emissivity 0.95. Some IR device such as thermal imager, thermal camera can adjust other emissivity. In this case, emissivity shall be set at 0.95 also.
– Follow user guide of Blackbody, set emissivity equal to 0.95, same with UUT.
– Base on UUT Distance to Spot ratio (D:S), use measuring tape to set the distance between Blackbody emitting surface and the UUT, that required for calibration process.
– Perform at least 5 points evenly distributed across the UUT range. Other test points may follow customer’s requirements.
– Begin at the lowest calibration point, example -10oC first, then increase for higher points.
– Set the Blackbody for the first calibration point and allow sufficient time to stabilize.
– Waiting for stable reading on UUT and the Blackbody, record repeat values.
– Repeat for remain test points.
– Calibration now completed, set and allow temperature of Blackbody near to normal room temperature.
– Turn off Blackbody and UUT, take out and secure all equipment.

6. Tolerance

– Tolerance given by the customer will be applied for all measurement, unless otherwise provisions.
– If the customer does not required, the specifications of manufacturer or other source of reference will be used.
– If not, the tolerance ±2oC or ±2%reading of UUT will be applied.
Nguyen Thuy

Share
Published by
Nguyen Thuy

Recent Posts

TECHMASTER’s New Headquarters: A Strategic Vision at a Prime Location – 175 Vo Nguyen Giap

New Techmaster Headquarters: Strategic Vision at a Prime Location In the dynamic context of Vietnam’s…

7 months ago

Techmaster is Keysight Technologies authorized distributor in Vietnam

Techmaster and Keysight Technologies Announce strategic partnership in Vietnam Techmaster is proud to announce its…

10 months ago

3M Scott SCBA: Breathe Safe in Any Situation

When working in toxic or low-oxygen environments, safety begins with proper respiratory protection. The 3M…

11 months ago

Vertical accuracy in oscilloscope

Vertical accuracy is a key factor that determines the measurement quality of an oscilloscope. It…

1 year ago

ISO/IEC 17025:2017 Scope Updates v015

In the latest update of the Scope of Accreditation to ISO/IEC 17025:2017 and ANSI/NCSL Z540-1-1994…

1 year ago

Vietnam Businesses: ISO 17025

ISO 17025 certification ensures that laboratories meet international standards for testing and calibration competence. It…

2 years ago