Techmaster introduce Calibration procedure for air flow meter

1. Scope of procedure

Unit Under Test Characteristics

Performance Range

Performance Method

Air flow rate

Up to 200 L/M

Comparison to standard flow meter

2. Standard and supporting device

– Standard Flow Meter ALICAT PCU, flow range up to 200L/min or equivalent
– Connecting tube with various size, fitting with in/out port of Unit Under Test (UUT) and standard meter (STD).
– Air flow generator range up to 200L/min or more, with flow regulator or control valve.

3. Environment conditions

– Temperature: (20 ÷ 25)oC

– Relative Humidity: (20 ÷ 65)%RH

Note: environment conditions above apply for Techmaster inhouse laboratories. Outside this bound conditions will be considered with customer for uncertainty and measurement results.

4. Preparation

– Select suitable tubes, connect Air Flow Generator – STD meter and UUT, at the suction side or arrow direction of each equipment.
– If UUT has control / adjustable valve, then connect UUT between Generator and STD meter.
– If UUT has internal pump, then Generator shall not be used. Connect UUT to STD meter directly.
– Teflon, quick-connector can be used if need.
– Ensure that system is no leak.
– Connect Air Flow Generator, STD meter, UUT to suitable power.
– Turn all on, allow warm up at least 30minutes or follow manufacturer’s instruction.

5. Perform

– Perform test points at: 25, 50, 75 and 100% UUT range, or follow customer’s requirement.
– Press measure button on UUT and STD meter, if needed.
– Slowly adjust flow rate output from the Generator, until STD meter reach nearly first test point.
– In case UUT has internal pump and flow setting function, then use UUT to generate test point.
– Allow for stable reading, record value of UUT and STD meter. Repeat 5 times.
– Repeat above steps for remain test points.
– Calibration now completed, set output air flow of Generator decreasing to zero.
– Turn off UUT and STD meter, take out of all components above.

6. Tolerance

– Device tolerance will be based on customer requirements, unless otherwise stated.
– If the customer has no requirements, it will be based on manufacturer specifications or other reference sources.
– If not, the tolerance ± (3% of reading +0.3%FS) will be applied.
Nguyen Thuy

Share
Published by
Nguyen Thuy

Recent Posts

TECHMASTER’s New Headquarters: A Strategic Vision at a Prime Location – 175 Vo Nguyen Giap

New Techmaster Headquarters: Strategic Vision at a Prime Location In the dynamic context of Vietnam’s…

2 months ago

Techmaster is Keysight Technologies authorized distributor in Vietnam

Techmaster and Keysight Technologies Announce strategic partnership in Vietnam Techmaster is proud to announce its…

6 months ago

3M Scott SCBA: Breathe Safe in Any Situation

When working in toxic or low-oxygen environments, safety begins with proper respiratory protection. The 3M…

6 months ago

Vertical accuracy in oscilloscope

Vertical accuracy is a key factor that determines the measurement quality of an oscilloscope. It…

10 months ago

ISO/IEC 17025:2017 Scope Updates v015

In the latest update of the Scope of Accreditation to ISO/IEC 17025:2017 and ANSI/NCSL Z540-1-1994…

1 year ago

Vietnam Businesses: ISO 17025

ISO 17025 certification ensures that laboratories meet international standards for testing and calibration competence. It…

1 year ago