facebook

Chauvin Arnoux MPC25: Multiparameter Tester

 

Key Features

  • Double display for easier reading
  • Manual switching of the parameters (pH/mV) and measurement ranges (conductivity)
  • Simultaneous measurement of pH or mV and conductivity
  • 2 separate analogue outputs
  • Compatible with Computer- Assisted Experimentation software

specifications Chauvin Arnoux MPC25: Multiparameter Tester

• 2000-point liquid crystal display

Ranges: pH: 0,00 to 14,00 pH; mV: -1999 to + 1999mV; Conductivity: 0-200mS/cm, 0-20mS/cm, 0-2000μS/cm, 0-200μS/cm

The voltage across the terminals of the conductivity cell: 100mV / 680Hz

Accuracy: ± 1%

Manual temperature compensation ( for the pH): 0 to 100°C

The temperature of use: 0 to 50°C

Maximum altitude: 2000m

Humidity: 5 to 80% at 35°C

Pollution class: Category 2

Category of use: 2

  • COMPLIANCE WITH STANDARDS

Safety: With the power supply provided: EN 61010-1. Normal environmental conditions (indoor use)

CEM: EN 61326, class B instrument, intermittent operation.

Consumption: 0,5W

Dimensions: 275 x 208 x 51 mm

Weight: 800g

The MPC 25 is provided with a power supply unit.

Power supply: 230V – 50/60Hz

Output: 9V

Introduction Chauvin Arnoux MPC25

Chauvin Arnoux MPC25 is a 2-in-1 instrument that is simultaneously a pH-meter and lab-stable conductivity meter. It is ideal for laboratories that need to perform pH and conductivity analyses.

For greater versatility, this instrument is equipped with two separate analog outputs (0-5 V) allowing it to be connected to one or more recording devices (flatbed plotter, logger).

These analog outputs also mean that the instrument can be integrated into Computer-AssistedExperimentation environments.

Features

– Accurately measure pH and conductivity in the laboratory

Make simple pH measurements, more detailed assessment of the acidity, alkalinity of the sample. Use BNC plug for pH/mV electrode with pH measurement range: 0.00 to 14.00 pH, electrode measurement ± 1999 mV. Conductivity measurement: 0 to 200 mS/cm in 4 ranges. Allow users to change manually. With just one device, users can simultaneously assess both pH and conductivity in solutions. Testing and evaluation in the laboratory become much faster and more efficient.

– Manual temperature compensation with a customizable range up to 100˚C

Because temperature can affect or change the results of the pH measurement. So the device has provided a manual temperature compensation mode. Users can adjust the temperature value with the Temperature dial on the device from 0 to 100°C. Helps maintain temperature balance. The thermal response for the buffer solutions ensures stable hydrogen ion activities.

As a result, Chauvin Arnoux MPC25 allows users to test the pH under different environmental conditions while ensuring reliability. This manual temperature compensation feature helps users to perform pH measurement requirements more efficiently and accurately. Meet the rigorous testing and evaluation requirements in the experiments.

– Manual pH calibration with 1 or 2 preset buffers pH4 and pH7

The Chauvin Arnoux MPC25 allows the user to perform single-point or two-point manual pH calibrations with two sets of buffers available. Ensures conditions for the highest accuracy of acid/base readings. Manual pH calibration. Help users make pH measurement simpler and easier. The data value after many measurements still has the reliability to support the analysis, measurement evaluation is also much more consistent and accurate.

– Design of 2 parallel displays – The counter can reach 2000 count

The design of the Chauvin Arnoux MPC25 desktop pH meter, helps users to see extremely clearly and intuitively with the design of 2 parallel screens that display pH / mV and conductivity parameters at the same time. Gives users an overview, general assessment, and measurements. The LCD display allows the user to observe the output data very clearly and intuitively through the analog display which can count up to 2000 counts. Support users to better process the data after measurement. Improved reliability of measurements. Meet different test requirements.

– Support 2 separate analog outputs (0-5V)

For added flexibility, this instrument is equipped with two separate analog outputs (0-5 V) allowing the meter to both record pH/mV and conductivity. And connect to one or more recording devices (flat plotters, loggers). With this analog output design, Chauvin Arnoux MPC25 allows users to integrate into a computer-aided test environment.

– Design compliant with NF EN61010-1 — NF EN 61326 (class B)

Chauvin Arnoux MPC25 meets all specified electrical equipment safety requirements and electrical equipment quality requirements for measurement, control, and laboratory use. Users are completely assured of pH measurement situations in a conductive solution environment. Ensure the highest level of safety. At the same time ensure to meet the most stringent testing requirements in the experiment.

Brand

Review Chauvin Arnoux MPC25: Multiparameter Tester

5 0% | 0 review
4 0% | 0 review
3 0% | 0 review
2 0% | 0 review
1 0% | 0 review
Review Chauvin Arnoux MPC25: Multiparameter Tester
Choose a image
0 character ( Minimum of 10)
    +

    There are no reviews yet.

    No comments yet

    RECENT PRODUCT

    hotline techmaster
    Hotline: 0908 173 345
    zalo techmaster Zalo: 0936 532 379 messenger techmaster Nhắn tin Facebook email techmaster Gửi Mail
    support techmaster