Calibration of torque measuring Instruments by the Snap-on TTC2000 system

In the current market, permissible error for torque calibration systems is commonly 1% of measured value. However, The Snap-On TTC2000 system has a low permissible error of 0.25% of the measured value.

Explore torque measuring tool calibration with Snap-On TTC2000 system with Techmaster (Calibration of torque measuring Instruments by the Snap-on TTC2000 system).

1. The calibration steps

  • Calibration of torque measuring Instruments by the Snap-on TTC2000 system

1.1 Calibration equipment

  • Standard equipment

– Snap-On TTC2000 Electronic Torque Calibration System

– The transducers corresponding to the measurement ranges are as follows: (0.45-5.65) N.m, (27.1-339) N.m, (81.3-813) N.m, (271-2710) N.m

  • The Measuring instruments to be verified

– The torque measuring tools within the calibration range of the SNAP-ON TTC2000 calibration system.

1.2 Calibration environmental conditions

+ Calibration conditions for Torque using the Snap-On TTC2000 system.

– Temperature: (23 ± 5) ºC, with a stability of ± 2 ºC during the calibration process.

– Humidity: ≤ 80% RH, with a stability of ± 5% RH during the calibration process.

1.3 Perform calibration

– Starting up the electronic torque calibration system

– Select a standard sensor with a suitable measurement range for the measuring instrument being checked.

1.3.1 Perform torque measurement verification within the ranges of (0.45-5.65) newton-met and (27.1-339) newton-met using the corresponding measuring instruments

– Attach instrument to calibration system per manufacturer’s instructions.

– Adjust the measurement unit of the calibration system to match the measurement unit of the measuring instrument. 

– Verify at 5 evenly spaced points as required by the customer, with 5 repetitions per point.

– Perform verification in both clockwise (CW) and counterclockwise (CCW) directions. If the measuring instrument is single-directional, verify only in that specific direction.

– Print the results after each measurement.

– Compare the average measured values from the calibration system with the set or displayed values of the torque measuring instrument.

1.3.2 Verify the measuring instrument within the range of (81.3-813) newton-met

– Replace the transducer with the corresponding measurement range mentioned above.

– Perform a similar verification following the same steps as mentioned above.

1.3.3 Verify the measuring instrument within the range of (81.3-813) newton-met

– Replace the transducer with the corresponding measurement range.

– Repeat the same verification process as mentioned above.

admin

Share
Published by
admin

Recent Posts

TECHMASTER’s New Headquarters: A Strategic Vision at a Prime Location – 175 Vo Nguyen Giap

New Techmaster Headquarters: Strategic Vision at a Prime Location In the dynamic context of Vietnam’s…

6 months ago

Techmaster is Keysight Technologies authorized distributor in Vietnam

Techmaster and Keysight Technologies Announce strategic partnership in Vietnam Techmaster is proud to announce its…

10 months ago

3M Scott SCBA: Breathe Safe in Any Situation

When working in toxic or low-oxygen environments, safety begins with proper respiratory protection. The 3M…

10 months ago

Vertical accuracy in oscilloscope

Vertical accuracy is a key factor that determines the measurement quality of an oscilloscope. It…

1 year ago

ISO/IEC 17025:2017 Scope Updates v015

In the latest update of the Scope of Accreditation to ISO/IEC 17025:2017 and ANSI/NCSL Z540-1-1994…

1 year ago

Vietnam Businesses: ISO 17025

ISO 17025 certification ensures that laboratories meet international standards for testing and calibration competence. It…

2 years ago